Abstract
ISO 17973:2002 specifies a method for calibrating the kinetic energy scales of Auger electron spectrometers with an uncertainty of 3 eV, for general analytical use in identifying elements at surfaces. In addition, it specifies a method for establishing a calibration schedule. It is applicable to instruments used in either direct or differential mode, where the resolution is less than or equal to 0,5 % and the modulation amplitude for the differential mode, if used, is 2 eV peak-to-peak. It is applicable to those spectrometers equipped with an inert gas ion gun or other method for sample cleaning and with an electron gun capable of operating at 4 keV or higher beam energy.
General information
-
Status: WithdrawnPublication date: 2002-10Stage: Withdrawal of International Standard [95.99]
-
Edition: 1Number of pages: 11
-
Technical Committee :ISO/TC 201/SC 7ICS :71.040.40
- RSS updates
Life cycle
-
Now
-
Revised by
WithdrawnISO 17973:2016